Code · CFR · Title 21 — Food and Drugs · Part 898 — Performance Standard for Electrode Lead Wires and Patient Cables · § 898.12

§ 898.12. Performance standard.

91 words·~1 min read·/us/cfr/t21/s§ 898.12·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Any connector in a cable or electrode lead wire having a conductive connection to a patient shall be constructed in such a manner as to comply with subclause 56.3(c) of the following standard: International Electrotechnical Commission
(IEC)601-1: Medical Electrical Equipment 601-1
(1988)Part 1: General requirements for safety Amendment No. 1
(1991)Amendment No. 2 (1995).
(b)Compliance with the standard shall be determined by inspection and by applying the test requirements and test methods of subclause 56.3(c) of the standard set forth in paragraph
(a)of this section.
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 898.12
Performance standard.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.