§ 886.5905. Oral electronic vision aid.

306 words·~1 min read·/us/cfr/t21/s§ 886.5905·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. An oral electronic vision aid is a battery-powered prescription device that contains an electrode stimulation array to generate electrotactile stimulation patterns that are derived from digital object images captured by a camera. It is intended to aid profoundly blind patients in orientation, mobility, and object recognition as an adjunctive device to other assistive methods such as a white cane or a guide dog.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Clinical performance testing must demonstrate an acceptable adverse event profile, including adverse events involving the mouth, tongue, and gums and demonstrate the effect of the stimulation to provide clinically meaningful outcomes. The clinical performance testing must also investigate the anticipated conditions of use, including potential use error, intended environment of use, and duration of use.
(2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including simulated moisture ingress, device durability, and battery reliability.
(3)Software verification, validation, and hazard analysis must be performed.
(4)Analysis/testing must validate electromagnetic compatibility.
(5)Analysis/testing must validate electrical safety.
(6)Analysis/testing must assess and validate wireless coexistence concerns.
(7)Any elements of the device that contact the patient must be demonstrated to be biocompatible.
(8)Training must include elements to ensure that the healthcare provider and user can identify the safe environments for device use, use all safety features of the device, and operate the device in the intended environment of use.
(9)Labeling for the trainer and user must include a summary of the clinical testing including adverse events encountered under use conditions, summary of study outcomes and endpoints, and information pertinent to use of the device including the conditions under which the device was studied (e.g., level of supervision or assistance, and environment of use). [80 FR 57092, Sept. 22, 2015]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 886.5905
Oral electronic vision aid.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.