§ 884.1710. Closed loop hysteroscopic insufflator with cutter-coagulator.

412 words·~2 min read·/us/cfr/t21/s§ 884.1710·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A closed loop hysteroscopic insufflator with cutter-coagulator is a prescription device configured for hysteroscopic insufflation, resection, and coagulation. It is used to perform diagnostic and surgical procedures (i.e., resection and coagulation). This device type contains a closed-loop recirculating fluid management system for the controlled delivery of filtered distension fluid. This device type also contains a bipolar radiofrequency device used in conjunction with a hysteroscope for resection and coagulation of intrauterine tissues.
(b)Classification. Class II (special controls). The special control(s) for this device are:
(1)The patient-contacting components of the device must be demonstrated to be biocompatible.
(2)Software validation, verification, and hazard analysis must be provided.
(3)Electrical equipment safety, including appropriate thermal and mechanical safety and electromagnetic compatibility
(EMC)testing must be performed.
(4)Device components that are labeled sterile must be validated to a sterility assurance level of 10−6.
(5)Shelf-life testing that demonstrates the device packaging maintains sterility and the functionality of the device is maintained following simulated shipping and handling must be provided to support the proposed shelf life.
(6)Non-clinical testing data must demonstrate the performance characteristics of the device. Detailed protocols and the test reports must be provided for each test.
(i)The following tests must be performed for the resection portion of the device:
(A)Mechanical testing to assess critical joint strength.
(B)Device electrode temperature testing.
(C)Coagulation depth testing.
(D)Simulated use testing.
(E)Device durability testing.
(ii)The following tests must be performed for the fluid management portion of the device:
(A)Mechanical testing to assess tensile strength of connections.
(B)Pressure testing that demonstrates the following parameters, including accuracy of the pressure displayed; appropriate detection and response to overpressure conditions; activation of a secondary overpressure relief valve at the maximum safe level; and all accessories within the fluid path meet the pressure requirements.
(C)Fluid delivery volume testing that demonstrates that the maximum fluid volume delivered is below a predefined level.
(D)Flow rate testing.
(E)Simulated use testing.
(F)Filtration testing.
(G)Blood filtration capacity testing.
(H)Tissue collection capacity testing.
(I)Filtrate characterization and testing that demonstrates that the continuous reintroduction of filtrate into the uterus does not pose a safety risk.
(7)Clinician labeling must include:
(i)Specific instructions and the clinical training needed for the safe use of the device.
(ii)Appropriate warnings, precautions, and information related to overpressurization.
(iii)Appropriate EMC information.
(iv)An expiration date/shelf life. [82 FR 35073, July 28, 2017]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 884.1710
Closed loop hysteroscopic insufflator with cutter-coagulator.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.