§ 882.5955. Temporary coil embolization assist device.

349 words·~2 min read·/us/cfr/t21/s§ 882.5955·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A temporary coil embolization assist device is a prescription device intended for temporary use in the neurovasculature to mechanically assist in the embolization of intracranial aneurysms with embolic coils. The device is delivered into the neurovasculature with an endovascular approach. This device is not intended to be permanently implanted and is removed from the body when the procedure is completed.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Clinical performance testing of the device must demonstrate the device performs as intended for temporary use as an endovascular device to assist in the coil embolization of intracranial aneurysms and must evaluate all adverse events, including tissue or vessel damage that could lead to dissection, perforation, hemorrhage, or vasospasm, thrombo-embolic events, and coil entanglement.
(2)The patient-contacting components of the device must be demonstrated to be biocompatible.
(3)Non-clinical performance testing must demonstrate the device performs as intended under anticipated conditions of use, including:
(i)Mechanical testing to demonstrate the device can withstand anticipated tensile, torsional, compressive, and tip deflection forces;
(ii)Mechanical testing to evaluate the radial forces exerted by the device;
(iii)Simulated use testing to demonstrate the device can be delivered to the target location in the neurovasculature and is compatible with embolic coils;
(iv)Dimensional verification testing;
(v)Radiopacity testing; and
(vi)Performance testing to evaluate the coating integrity and particulates under simulated use conditions.
(4)Animal testing under anticipated use conditions must evaluate all adverse events, including damage to vessels or tissues.
(5)Performance data must support the sterility and pyrogenicity of the device.
(6)Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
(7)The labeling must include:
(i)Instructions for use;
(ii)A detailed summary of the device technical parameters, including compatible delivery catheter dimensions and device sizing information;
(iii)A summary of the clinical testing results, including a detailed summary of the device- and procedure-related complications and adverse events; and
(iv)A shelf life. [86 FR 70733, Dec. 13, 2021]
Connections1 cite this
Citation graph
cites case law
§ 882.5955
Temporary coil embolization assist device.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.