§ 882.1630. Cranial motion measurement device.

242 words·~1 min read·/us/cfr/t21/s§ 882.1630·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A cranial motion measurement device is a prescription device that utilizes accelerometers to measure the motion or acceleration of the skull. These measurements are not to be used for diagnostic purposes.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
(i)Hardware specifications must be provided. Additionally, verification and validation testing as well as a hazard analysis must be performed.
(ii)Software must be described in detail in the Software Requirements Specification
(SRS)and Software Design Specification (SDS). Additionally, software verification and validation testing as well as a hazard analysis must be performed.
(2)The device parts that contact the patient must be demonstrated to be biocompatible.
(3)The device must be designed and tested for electrical, thermal, and mechanical safety, and electromagnetic compatibility (EMC).
(4)Clinical performance testing must demonstrate the accuracy, precision, stability, and repeatability of measuring cranial motion per the intended use in the intended use environment.
(5)The labeling must include:
(i)The intended use population and the intended use environment.
(ii)Instructions for technicians to convey to patients regarding the collection of cranial acceleration data to ensure device measurement accuracy, precision, stability, and repeatability.
(iii)Information allowing clinicians to understand potential sources of variability in the measurement to help recognize and identify changes in the measurement. [82 FR 35071, July 28, 2017]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 882.1630
Cranial motion measurement device.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.