Tap any paragraph to write a margin note. Your notes collect in the Desk below the text and file under cases with @. The side-by-side margin rail opens on a larger screen.

Code · CFR · Title 21 — Food and Drugs · Part 878 — General and Plastic Surgery Devices · § 878.4783

§ 878.4783. Negative pressure wound therapy device for reduction of wound complications.

400 words·~2 min read·/us/cfr/t21/s§ 878.4783·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A negative pressure wound therapy device for reduction of wound complications is a powered suction pump intended for wound management and reduction of wound complications via application of negative pressure to the wound, which removes fluids, including wound exudate, irrigation fluids, and infectious materials. This device type is intended for use with wound dressings classified under § 878.4780. This classification does not include devices intended for organ space wounds.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Clinical data must demonstrate that the device performs as intended under anticipated conditions of use and evaluate the following:
(i)Wound complication rates; and
(ii)All adverse events.
(2)The patient-contacting components of the device must be demonstrated to be biocompatible.
(3)Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4)Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
(5)Usability testing must demonstrate that intended users can correctly use the device, based solely on reading the instructions for use.
(6)Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested in a worst-case scenario for the intended use life:
(i)Ability to maintain pressure levels at the wound site under a worst-case scenario for the intended use life;
(ii)Fluid removal rate consistent with the wound types specified in the indications for use; and
(iii)Timely triggering of all alarms.
(7)Performance data must demonstrate the electrical safety and electromagnetic compatibility
(EMC)of the device.
(8)Software verification, validation, and hazard analysis must be performed.
(9)Labeling must include the following:
(i)Instructions for use;
(ii)A summary of the device technical specifications, including pressure settings, modes (e.g., continuous or intermittent), alarms, and safety features;
(iii)Compatible components and devices;
(iv)A summary of the clinical evidence for the indications for use;
(v)A shelf life for sterile components; and
(vi)Use life and intended use environments.
(10)For devices intended for use outside of a healthcare facility, patient labeling must include the following:
(i)Information on how to operate the device and its components and the typical course of treatment;
(ii)Information on when to contact a healthcare professional; and
(iii)Use life. [86 FR 70734, Dec. 10, 2021]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 878.4783
Negative pressure wound therapy device for reduction of wound complications.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.