Code · CFR · Title 21 — Food and Drugs · Part 878 — General and Plastic Surgery Devices · § 878.4452

§ 878.4452. Nonabsorbable expandable hemostatic sponge for temporary internal use.

411 words·~2 min read·/us/cfr/t21/s§ 878.4452·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A nonabsorbable expandable hemostatic sponge for temporary internal use is a prescription device intended to be placed temporarily into junctional, non-compressible wounds, which are not amenable to tourniquet use, to control bleeding until surgical care is acquired. The sponges expand upon contact with blood to fill the wound cavity and provide a physical barrier and pressure that facilitates formation of a clot. The device consists of sterile, nonabsorbable radiopaque compressed sponges and may include an applicator to facilitate delivery into a wound.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Performance data must demonstrate the biocompatibility of patient-contacting components.
(2)Performance data must demonstrate the sterility of patient-contacting components including endotoxin and pyrogenicity assessments.
(3)Performance data must support device stability by demonstrating continued sterility of the patient-contacting components of the device, package integrity, and device functionality over the requested shelf life.
(4)Assessment of material characteristics must be sufficient to support safety under anticipated conditions of use. Assessments must include the following:
(i)Material specifications.
(ii)Immunogenicity.
(iii)Viral inactivation for animal-derived materials.
(5)Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i)Absorption capacity.
(ii)Extent of swelling.
(iii)Mechanical properties.
(iv)Expansion force/pressure.
(v)Radiopacity.
(vi)Deployment/applicator functionality.
(6)In vivo performance data must demonstrate safe and effective use by verifying that the device performs as intended under anticipated conditions of use. Appropriate analysis/testing must demonstrate that the product: Controls bleeding, does not promote adverse local or systemic effects, and can be completely removed from the wound. The following performance characteristics must be tested:
(i)Deployment.
(ii)Control of bleeding.
(iii)Radiopacity.
(iv)Retrieval.
(v)Assessment of local and systemic effects.
(7)Human factors testing and analysis must validate that the device design and labeling are sufficient for appropriate use by emergency responders deploying the device as well as surgeons retrieving the device from wounds.
(8)Labeling must include:
(i)Specific instructions for deployment by emergency responders and retrieval by surgeons.
(ii)Warnings, cautions, and limitations needed for safe use of the device.
(iii)Information on how the device operates and the typical course of treatment.
(iv)A detailed summary of the in vivo and human factors testing pertinent to use of the device.
(v)Appropriate imaging information to ensure complete retrieval of device.
(vi)An expiration date/shelf life. [79 FR 34224, June 16, 2014]
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.