§ 878.4420. Electrosurgical device for over-the-counter aesthetic use.

278 words·~1 min read·/us/cfr/t21/s§ 878.4420·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. An electrosurgical device for over-the-counter aesthetic use is a device using radiofrequency energy to produce localized heating within tissues for non-invasive aesthetic use.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested: Over-heating, power accuracy radiofrequency, pulse cycle, waveform, pulse duration, and device characterization parameters.
(2)Label comprehension and self-selection performance evaluation must demonstrate that the intended over-the-counter users can understand the package labeling and correctly choose the device for the indicated aesthetic use.
(3)Usability performance evaluation must demonstrate that the over-the-counter user can correctly use the device, based solely on reading the directions for use, to treat the indicated aesthetic use.
(4)Clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use to achieve the intended aesthetic results.
(5)The patient-contacting components of the device must be demonstrated to be biocompatible.
(6)Instructions for cleaning the device must be validated.
(7)Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety, including the mechanical integrity, of the device.
(8)Software verification, validation, and hazard analysis must be performed.
(9)Labeling must include:
(i)Warnings, precautions, and contraindications to ensure the safe use of the device for the over-the-counter users.
(ii)A statement that the safety and effectiveness of the device's use for uses other than the indicated aesthetic use are not known.
(iii)A summary of the clinical information used to establish effectiveness for each indicated aesthetic usage and observed adverse events. [81 FR 42244, June 29, 2016]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 878.4420
Electrosurgical device for over-the-counter aesthetic use.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.