§ 878.4371. Irrigating wound retractor device.

279 words·~1 min read·/us/cfr/t21/s§ 878.4371·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. An irrigating wound retractor device is a prescription device intended to be used by a surgeon to retract the surgical incision, to provide access to the surgical wound, to protect and irrigate the surgical wound, and to serve as a conduit for removal of fluid from the surgical wound.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)The patient-contacting components of the device must be demonstrated to be biocompatible and evaluated for particulate matter.
(2)Performance data must demonstrate the sterility and pyrogenicity of the patient-contacting components of the device.
(3)Performance data must support shelf life by demonstrating continued functionality and sterility of the device over the identified shelf life.
(4)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must:
(i)Characterize the tear resistance, tensile strength, and elongation properties of the barrier material;
(ii)Demonstrate that the liquid barrier material is resistant to penetration by blood, and is non-flammable;
(iii)Characterize the forces required to deploy the device;
(iv)Characterize the device's ranges of operation, including flow rates and maximum suction pressures;
(v)Demonstrate the ability of the device irrigation apparatus to maintain a user defined or preset flow rate to the surgical wound; and
(vi)Demonstrate the ability of the device to maintain user defined or preset removal rates of fluid from the surgical wound.
(5)The labeling must include or state the following information:
(i)Device size or incision length range;
(ii)Method of sterilization;
(iii)Flammability classification;
(iv)Non-pyrogenic;
(v)Shelf life; and
(vi)Maximum flow rate and suction pressure. [83 FR 24, Jan. 2, 2018]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 878.4371
Irrigating wound retractor device.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.