§ 876.5011. Metallic biliary stent system for benign strictures.

374 words·~2 min read·/us/cfr/t21/s§ 876.5011·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A metallic biliary stent system for benign strictures is a prescription device intended for the treatment of benign biliary strictures. The biliary stents are intended to be left indwelling for a limited amount of time and subsequently removed. The device consists of a metallic stent and a delivery system intended to place the biliary stent in the bile duct. This device type is not intended for use in the vasculature.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Clinical performance testing must demonstrate or provide the following:
(i)The ability to safely place and subsequently remove the stent after the maximum labeled indwell period.
(ii)All adverse event data including bile duct obstruction and trauma to the bile duct.
(iii)The stent resolves strictures during the maximum labeled indwell period.
(iv)Stricture resolution is maintained post-stent removal.
(2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
(i)Corrosion testing to demonstrate that the stent maintains its integrity during indwell and does not release potentially toxic levels of leachables.
(ii)Stent dimensional testing supports the intended use.
(iii)Compression and expansion forces must be characterized.
(iv)The delivery catheter must deliver the stent to the intended location and the stent must not be adversely impacted by the delivery catheter during deployment and catheter withdrawal.
(v)The delivery system must withstand clinically anticipated forces.
(vi)Compatibility in a magnetic resonance environment.
(3)All patient contacting components of the device must be demonstrated to be biocompatible.
(4)Performance data must demonstrate the sterility of the device components intended to be provided sterile.
(5)Shelf life testing must demonstrate that the device maintains its performance characteristics and that packaging maintains sterility for the duration of the labeled shelf life.
(6)Labeling for the device must include:
(i)A detailed summary of the clinical testing including device effectiveness, and device- and procedure-related adverse events.
(ii)Appropriate warning(s) to accurately ensure usage of the device for the intended patient population.
(iii)Shelf life.
(iv)Compatibility information for use in the magnetic resonance environment.
(v)Stent foreshortening information supported by dimensional testing. [81 FR 45231, July 13, 2016]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 876.5011
Metallic biliary stent system for benign strictures.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.