§ 876.4350. Fluid jet system for prostate tissue removal.

343 words·~2 min read·/us/cfr/t21/s§ 876.4350·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A fluid jet system for prostate tissue removal is a prescription device intended for the resection and removal of prostatic tissue for the treatment of benign prostatic hyperplasia. The device cuts tissue by using a pressurized jet of fluid delivered to the prostatic urethra. The device is able to image the treatment area, or pairs with an imaging modality, to monitor treatment progress.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Clinical performance testing must evaluate the following:
(i)All adverse events associated with the device, and
(ii)Improvement in lower urinary tract symptoms (LUTS).
(2)Physician training must be provided that includes:
(i)Information on key aspects and use of the device, and
(ii)Information on how to override or stop resection.
(3)Animal testing must demonstrate that the device resects targeted tissue in a controlled manner without injury to adjacent non-target tissues.
(4)Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i)Measurement of targeting accuracy and reproducibility of high velocity fluid jet, and
(ii)High pressure fluid jet verification testing at target and non-target tissues.
(5)Software verification, validation, and hazard analysis must be performed.
(6)The patient-contacting elements of the device must be demonstrated to be biocompatible.
(7)Performance data must demonstrate the electrical safety and electromagnetic compatibility of the device.
(8)Performance data must demonstrate the sterility of the patient-contacting components of the device.
(9)Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(10)Performance data must validate the instructions for reprocessing and reliability of reusable components.
(11)Labeling must include the following:
(i)A section that summarizes the clinical testing results, including the adverse event profile and improvement in LUTS;
(ii)A shelf life for single use components;
(iii)A use life for reusable components; and
(iv)Reprocessing instructions for reusable components. [83 FR 27897, June 15, 2018]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 876.4350
Fluid jet system for prostate tissue removal.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.