§ 868.5273. Positive airway pressure delivery system.

304 words·~1 min read·/us/cfr/t21/s§ 868.5273·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. A positive airway pressure delivery system is a prescription noninvasive ventilatory device that delivers expiratory positive airway pressure for patients suffering from obstructive sleep apnea. The system also provides positive airway pressure during incipient apnea. The system may include a dedicated flow generator and a patient interface.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)The patient-contacting components of the device must be demonstrated to be biocompatible.
(2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following:
(i)Waveform testing must simulate breathing conditions and evaluate pressure and airflow response over a range and combination of high and low breath rates and tidal volumes.
(ii)Use life testing must demonstrate adequate device performance over the labeled use life of the device.
(iii)Device integrity testing must demonstrate that the device can withstand typical forces expected during use.
(iv)Carbon dioxide rebreathing testing must be performed.
(v)System flow rate, maximum expiratory pressure, inhalation pressure, and intra-mask static pressure testing must be performed.
(vi)Air bolus testing must demonstrate that the device can withstand worst-case scenario air pressures.
(vii)Maximum limited pressure testing of the flow generator in single fault condition must be performed.
(viii)Maximum output temperature testing of delivered gas, if humidified, must be performed.
(3)Performance data must validate reprocessing instructions for any reusable components of the device.
(4)Performance data must demonstrate the electrical, thermal, and mechanical safety and the electromagnetic compatibility of the device.
(5)Software verification, validation, and hazard analysis must be performed.
(6)Labeling must include the following:
(i)Therapy pressure range;
(ii)Use life and replacement schedule for all components;
(iii)Cleaning instructions; and
(iv)Instructions for assembly and connection of device components. [83 FR 52966, Oct. 19, 2018]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 868.5273
Positive airway pressure delivery system.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.