Code · CFR · Title 21 — Food and Drugs · Part 868 — Anesthesiology Devices · § 868.5105

§ 868.5105. External negative pressure airway aid.

312 words·~1 min read·/us/cfr/t21/s§ 868.5105·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Identification. An external negative pressure airway aid is a prescription device that applies negative pressure to a patient's neck to aid in providing a patent airway during procedures requiring anesthesia.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Clinical performance testing must document any adverse events observed during clinical use, including impaired blood flow, and demonstrate that the device performs as intended under anticipated conditions.
(2)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated patient positions, does not fail during use, and does not lose negative pressure capability. The following testing should be performed:
(i)Ability of the device to maintain a seal during various patient positions;
(ii)Device leakage testing to demonstrate the device maintains vacuum;
(iii)Drop testing to ensure the device does not incur functional damage after dropping the device; and
(iv)Functional testing after high and low storage temperature.
(3)All patient contacting components must be demonstrated to be biocompatible.
(4)Labeling must include:
(i)A summary of clinical testing results, including any adverse events and evidence that effectiveness has been achieved.
(ii)Technical specifications of the device, including collar sizes, maximum duration of use, operating temperature, and storage temperature range.
(iii)Technical specifications of the vacuum source, including maximum vacuum level and operational vacuum level.
(iv)Instructions for use that includes how to place the device, determination of size, verification of suction, reference to training materials, and information on troubleshooting the device if it does not attach properly.
(v)A warning to screen patients for carotid artery disease due to the probable risk of the device to dislodge arterial plaques in the carotid artery.
(vi)A warning to exclude patients with anatomical abnormalities.
(vii)A warning not to use the device during medical procedures involving medications that contain propofol. [82 FR 60867, Dec. 26, 2017]
Connections1 cite this
Cited by 1 section
Citation graph
cites case law
§ 868.5105
External negative pressure airway aid.
Fed. Reg.×1
Cites 0Cited by 1 across 1 source
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.