§ 606.60. Equipment.

347 words·~2 min read·/us/cfr/t21/s§ 606.60·

A research copy — for the controlling text, always check the official state or federal source. Not legal advice.

(a)Equipment used in the collection, processing, compatibility testing, storage and distribution of blood and blood components shall be maintained in a clean and orderly manner and located so as to facilitate cleaning and maintenance. The equipment shall be observed, standardized and calibrated on a regularly scheduled basis as prescribed in the Standard Operating Procedures Manual and shall perform in the manner for which it was designed so as to assure compliance with the official requirements prescribed in this chapter for blood and blood products.
(b)Equipment that shall be observed, standardized and calibrated with at least the following frequency, include but are not limited to: Equipment Performance check Frequency Frequency of calibration Temperature recorderCompare against thermometerDailyAs necessary. Refrigerated centrifugeObserve speed and temperatureEach day of use Do. Hematocrit centrifugeStandardize before initial use, after repairs or adjustments, and annually. Timer every 3 mo. General lab centrifugeTachometer every 6 mo. Automated blood-typing machineObserve controls for correct resultsEach day of use HemoglobinometerStandardize against cyanmethemoglobin standard......do RefractometerStandardize against distilled water......do Blood container scaleStandardize against container of known weight......doAs necessary. Water bathObserve temperature......do Do. Rh view box......do......do Do. Autoclave......doEach time of use Do. Serologic rotatorsObserve controls for correct resultsEach day of useSpeed as necessary. Laboratory thermometersBefore initial use. Electronic thermometersMonthly. Vacuum blood agitatorObserve weight of the first container of blood filled for correct resultsEach day of useStandardize with container of known mass or volume before initial use, and after repairs or adjustments.
(c)Equipment employed in the sterilization of materials used in blood collection or for disposition of contaminated products shall be designed, maintained and utilized to ensure the destruction of contaminating microorganisms. The effectiveness of the sterilization procedure shall be no less than that achieved by an attained temperature of 121.5 °C (251 °F) maintained for 20 minutes by saturated steam or by an attained temperature of 170 °C (338 °F) maintained for 2 hours with dry heat. [40 FR 53532, Nov. 18, 1975; 40 FR 55849, Dec. 2, 1975, as amended at 45 FR 9261, Feb. 12, 1980; 57 FR 11263, Apr. 2, 1992; 57 FR 12862, Apr. 13, 1992]
★   the supreme law of the land   ★
Don't Tread on Me
E Pluribus Unum — out of many, one

"If you don't know your rights, you don't have any."

Marginalia · a citizen's law index
A research desk, not legal advice. Always read the cited source before relying on a summary.
Questions or an issue? support@self-law.org
disclaimerMarginalia is a research index, not a law firm. Nothing on this site is legal, tax, or financial advice and no attorney–client relationship is formed by using it. Statutes, regulations, and case law change; summaries, search results, AI output, and member posts may be incomplete, out of date, or wrong. Any interpretation drawn from material on this site should be validated by a licensed attorney in your jurisdiction before you act on it.